R+D+i

Torque test of the External Hexagon

To confirm the quality of the MPI Privilege material and implant design, the company has conducted a study to demonstrate the high resistance to breakage of the external hexagon....

Hydrophilic Properties of the Mpi Privilege Implant System
The test results show that the tension of the implant surface exhibit a hydrophilic behaviour both on the treated surface as on the machined surface. In literature hydrophilic surfaces are taken...
IPCM - Static Resistance to Compression Bending

The results of the static tests carried out on the MPI Privilege implants of “Internal Connection” are satisfactory, due to the fact that they have supported higher static load values than usual...

IPCM - Fatigue Test

The fatigue limit of the ”MPI Privilege Implant of Internal Connection” has reached a value of 362.4 N, with the geometric arrangement shown in ISO 14801:2007....

IPHE - Static Resistance to Compression Bending

In Static Resistance tests, the Implant of External Universal Hexagon has reached a value at the yield strength of 521 N. The maximum bite force with incisors varies between 50 and 370 N....

IPHE - Fatigue Test

The fatigue limit of the ”MPI Privilege Implant of External Connection” has obtained a value of 204 N, respecting the geometric arrangement shown in ISO 14801:2007.

Due to...